The current range of the dc ammeter may be further
extended by a number of shunts, selected by a range switch. Such meter is
called a multi range ammeter.
The multi range ammeter as shown in below figure.
The circuit has four shunts,R1, R2,
R3 and R4 which can be placed in parallel with the
movement to give four different current ranges I1,,I2,,I3
and I4.
R1=Im
Rm / I1-Im for current range I1;
R2= Im Rm/ I2-Im
for current range I2;
R3= Im Rm/ I3-Im
for current range I3;
R4= Im Rm/ I4-Im
for current range I4.
Switch “S” is a multi position switch, make before
break type switch. It is used for range changing. This switch protects the
meter movement from being damaged without a shunt during range changing.
Multi range ammeters are used for ranges up to 50A.
When using a multi range ammeter, first use the highest current range, then
decrease the range until good upscale reading is obtained. The resistance used
for the various a ranges are of very high precision values, hence the cost of
the meter increases.